Calibration of Flow Cytometers Enables Reproducible Measurements of Extracellular Vesicle Concentrations and Reference Range Establishment

Bettin, Britta A. ✉; Li, Bo; Falkena, Kim; Van Leeuwen, Ton G.; Gollwitzer, Christian; Varga, Zoltan [Varga, Zoltán (biofizika, kolloi...), szerző] Fizikai Kémia és Anyagtudományi Tanszék (BME / VBK); Anyag- és Környezetkémiai Intézet (HRN TTK); Biológiai Nanokémiai Kutatócsoport (HRN TTK / AKI); Ajzenberg, Nadine; Antovic, Jovan P.; Berckmans, Pascale; Buzas, Edit I. [Buzás, Edit Irén (Immunbiológia), szerző] Genetikai, Sejt- és Immunbiológiai Intézet (SE / AOK / I); HCEMM-SE Extracelluláris Vezikula Kutatócsoport (SE / AOK / I / GSII); HUN-REN-SE Transzlációs Extracelluláris Vezikul... (SE / AOK / I / GSII); Carney, Randy P.; Cook, Sean; Dignat-george, Francoise; Faille, Dorothee; Giebel, Bernd; Jones, Jennifer C.; Kim, Yohan; Lacroix, Romaric; Lannigan, Joanne; Lucien, Fabrice; Maaninka, Katariina; De Menezes, Erika G. Marques; Meyer, Annette; Mizenko, Rachel R.; Nelissen, Inge; Nolan, John; Norris, Philip J.; Pink, Desmond; Poudel, Sumeet; Robert, Stephane; Siljander, Pia R. -m.; Tang, Vera A.; Tertel, Tobias; Broeck, Tina Van Den; Wang, Lili; Welsh, Joshua A.; Nieuwland, Rienk; Van Der Pol, Edwin

Angol nyelvű Sokszerzős vagy csoportos szerzőségű szakcikk (Folyóiratcikk) Tudományos
Megjelent: JOURNAL OF EXTRACELLULAR VESICLES 2001-3078 14 (12) Paper: e70189 , 18 p. 2025
  • SJR Scopus - Cell Biology: D1
Szakterületek:
  • Biológiai tudományok
The concentration of cells is a key component of modern blood tests. Given the biomarker potential of extracellular vesicles (EVs) in blood, we aimed to establish reference ranges for blood cell-derived EVs using flow cytometry. To address the orders-of-magnitude variability in reported EV concentrations between different flow cytometers (FCMs), we first validated a calibration methodology to enable reproducible EV concentration measurements. The methodology was evaluated in an interlaboratory comparison study and shows that calibration reduces the median absolute deviation of EV concentrations measured on 25 different FCMs from 67 % to 25 %–31 %. The calibration methodology was then used to determine reference ranges of erythrocyte-, leukocyte-, and platelet-derived EVs in human blood plasma in a cohort of healthy individuals (n = 224). This study demonstrates that calibration enables comparable concentration measurements of blood cell-derived EVs, thereby bringing EVs one step closer to clinical applications.
Hivatkozás stílusok: IEEEACMAPAChicagoHarvardCSLMásolásNyomtatás
2026-09-06 05:34