Background: Recurrences due to discontinuity in ablation lines are substantial after
pulmonary vein isolation (PVI) with radiofrequency ablation for atrial fibrillation.
Data are scarce regarding the durability predictors for very high-power short-duration
(vHPSD, 90 W/4 s) ablation. Methods: A total of 20 patients were enrolled, who underwent
90 W PVI and a mandatory remapping procedure at 3 months. First-pass isolation (FPI)
gaps, and acute pulmonary vein reconnection (PVR) sites were identified at the index
procedure; and chronic PVR sites were identified at the repeated procedure. We analyzed
parameters of ablation points (n = 1357), and evaluated their roles in predicting
a composite endpoint of FPI gaps, acute and chronic PVR. Results: In total, 45 initial
ablation points corresponding to gaps in the ablation lines were analyzed. Parameters
associated with gaps were interlesion distance (ILD), baseline generator impedance,
mean current, total charge, and loss of catheter–tissue contact. The optimal ILD cut-off
for predicting gaps was 3.5 mm anteriorly, and 4 mm posteriorly. Conclusions: Biophysical
characteristics dependent on generator impedance could affect the efficacy of vHPSD
PVI. The use of smaller ILDs is required for effective and durable PVI with vHPSD
compared to the consensus targets with lower power ablation, and lower ILDs for anterior
applications seem necessary compared to posterior points.