Dose indices of kilovoltage cone beam computed tomography for various image guided radiotherapy protocols.

Gazdag-Hegyesi, Szilvia ✉ [Gazdag-Hegyesi, Szilvia (Orvosfizika), author] National Institute of Oncology; Gáldi, Ádám [Gáldi, Ádám (Orvos fizika), author] National Institute of Oncology; Major, Tibor [Major, Tibor (klinikai sugárfizika), author] National Institute of Oncology; Onkológiai Tanszék (SU / FM / C); Pesznyák, Csilla [Pesznyák, Csilla (Orvosfizika), author] National Institute of Oncology; Institute of Nuclear Technology (BUTE / FNS)

English Article (Journal Article) Scientific
Published: RADIATION PROTECTION DOSIMETRY 0144-8420 1742-3406 199 (8-9) pp. 983-988 2023
  • SJR Scopus - Medicine (miscellaneous): Q3
Identifiers
Subjects:
  • Radiology, nuclear medicine and medical imaging
Varian Halcyon (HA) linear accelerators (LINAC) require kilovoltage cone-beam computed tomography (CT) acquisition for all patients before all treatment fractions. The aim of this study is to compare dose indices of various available protocols with different calculation and measurement methods. CT dose index (CTDI) in mGy is a measure of the radiation dose output of a CT scanner. A pencil ionisation chamber was used to measure dose index free air and in a standard CTDI phantom for different imaging protocols of HA and TrueBeam LINACs. For the point measurements, we found large deviations between displayed and calculated low CTDI values: 26.6 and 27.1% at Head low dose and Breast protocol, respectively. The calculated values were always larger than the displayed ones for all protocols and measurement setups. For the point measurements similar results were found as in the international literature, where the displayed values the measured CTDIs.
Citation styles: IEEEACMAPAChicagoHarvardCSLCopyPrint
2025-04-26 11:35