Implant for Augmentation of Cerebral Blood Flow Trial-1 (ImpACT-1). A single-arm feasibility
study evaluating the safety and potential benefit of the Ischemic Stroke System for
treatment of acute ischemic stroke
BackgroundThe Ischemic Stroke System is a novel device designed to deliver stimulation
to the sphenopalatine ganglion(SPG). The SPG sends parasympathetic innervations to
the anterior cerebral circulation. In rat stroke models, SPG stimulation results in
increased cerebral blood flow, reduced infarct volume, protects the blood brain barrier,
and improved neurological outcome. We present here the results of a prospective, multinational,
single-arm, feasibility study designed to assess the safety, tolerability, and potential
benefit of SPG stimulation inpatients with acute ischemic stroke(AIS).MethodsPatients
with anterior AIS, baseline NIHSS 7-20 and ability to initiate treatment within 24h
from stroke onset, were implanted and treated with the SPG stimulation. Patients were
followed up for 90 days. Effect was assessed by comparing the patient outcome to a
matched population from the NINDS rt-PA trial placebo patients.ResultsNinety-eight
patients were enrolled (mean age 57years, mean baseline NIHSS 12 and mean treatment
time from stroke onset 19h). The observed mortality rate(12.2%), serious adverse events
(SAE) incidence(23.5%) and nature of SAE were within the expected range for the population.
The modified intention to treat cohort consisted of 84 patients who were compared
to matched patients from the NINDS placebo arm. Patients treated with SPG stimulation
had an average mRS lower by 0.76 than the historical controls(CMH test p = 0.001).ConclusionThe
implantation procedure and the SPG stimulation, initiated within 24hr from stroke
onset, are feasible, safe, and tolerable.